Keithley Model 2000 - detected, S/N ???????, Version: A20 /A02 Environment temperature = 31.87 C, humidity = 35.60 RH, pressure = 1005.5598 hPa HP 3458A detected TEMP? = 38.1 C Ambient temp = +24.7C HP 3458A DCV -7.834272296E-07 VDC, K2000 DUT 3.078535757E-06 VDC, Delta -1254480.471 ppm MFC error = 1238.234 ppm HP 3458A DCV 9.999855743E-03 VDC, K2000 DUT 1.000182233E-02 VDC, Delta -196.623 ppm MFC error = 60.005 ppm HP 3458A DCV 1.000006746E-01 VDC, K2000 DUT 1.000024920E-01 VDC, Delta -18.174 ppm MFC error = -17.032 ppm HP 3458A DCV 9.999989351E-01 VDC, K2000 DUT 1.000001607E+00 VDC, Delta -2.672 ppm MFC error = -4.894 ppm HP 3458A DCV 9.999971088E+00 VDC, K2000 DUT 1.000000503E+01 VDC, Delta -3.394 ppm MFC error = 12.827 ppm HP 3458A DCV 1.000000769E+02 VDC, K2000 DUT 9.999921813E+01 VDC, Delta 8.588 ppm MFC error = 21.998 ppm HP 3458A DCV 1.000003027E+03 VDC, K2000 DUT 1.000014860E+03 VDC, Delta -11.833 ppm MFC error = 142.410 ppm Measured REF DCI 5 : 9.999802284E-05 ADC, stored DCI = 0.099986 MFC error = 4.031 ppm Measured REF DCI 5 : 9.999978065E-04 ADC, stored DCI = 0.999996 MFC error = 9.888 ppm Measured REF DCI 5 : 9.999997661E-03 ADC, stored DCI = 9.999901 MFC error = -12.991 ppm Measured REF DCI 5 : 9.999987479E-02 ADC, stored DCI = 100.001299 MFC error = -78.172 ppm Measured REF DCI 5 : 9.999943627E-01 ADC, stored DCI = 1.000078 MFC error = 19.204 ppm MFC error = -310.981 ppm MFC error = -312.800 ppm MFC error = -57.848 ppm MFC error = -20.603 ppm MFC error = 78.645 ppm MFC error = 1045.179 ppm Calibrations count = 3 Calibration last date = 2016,07,27 :CAL:PROT:CODE 'KI002000' Error status = -113,"Undefined header" Error status = 0,"No error" Error status = 0,"No error" CAL MFC error = -4.150 ppm :CAL:PROT:DC:STEP3 9.999978 Error status = 0,"No error" CAL MFC error = -18.864 ppm :CAL:PROT:DC:STEP4 -9.999971 Error status = 0,"No error" CAL MFC error = 4.726 ppm :CAL:PROT:DC:STEP5 100.001143 Error status = 404,"100 vdc full scale error"